 |
Why should the insulation on surgical
electrode probes be tested?
|
 |
It has been determined that insulation failure of
active electrode surgical probes (bipolar and monopolar)
is one of the major reasons why patients receive electrosurgical
burns to unintended tissues. Even the smallest hole
in the insulation of the probe will endanger the patient's
safety since the escaping electrical energy will usually
cause burns to the patient outside of the surgeon's
view. Visual inspection of the integrity of the probe's
insulation will often miss the small holes and cracks
that occur.
|
 |
What are the causes of Pinholes?
|
 |
While much care is taken in the handling of these
probes they are still subjected to mechanical stress
due to the sterilization methods or cleaning agents
used, such as alcohol or glutaraldehyde.
- These mechanical stresses can produce pinholes, nicks
or cracks in the protective insulation of the probe.
- Handling of the electrode surgical probe during the
surgery may also cause damage to the insulation.
- There is also a possibility of defects caused to the
probe during packaging or that may occur to it during
or after being shipped from the manufacturer. |
 |
Can any other products be tested using the Micromed
PD-8K?
|
 |
Yes, any conductive element with coating can be tested
providing that the conductive element is easily accessible.
For example, EKG, EEG probes. |
 |
What should be done to ensure the insulation's integrity?
|
 |
A vigorous program of regular inspection of probes'
insulation should be practiced. Formal procedures should
be set-up to minimize the potential damage to the probe's
insulation. This procedure should incorporate the probe's
insulation testing on a regular basis. On of the best
and most cost effective way to test the probe's insulation's
integrity is to utilize the MicroMed PD-8K detection
unit. It is suggested that the probe be tested post-surgery
in the Sterile Department after the probe has been cleaned
and just prior to sterilization. |
 |
What is the overall Benefit?
|
 |
With 10's of thousands of unwanted electrosurgical
burns occurring every year it is essential to reduce
this number to protect the hospital and the physician
from liability issues but more importantly to protect
the patient from the numerous side effects from these
types of burns and the associated costs to the hospital
for the extra care required for these cases. |
 |
On what principle does the Micromed Detector operate?
|
 |
The unit is a low frequency high voltage generator
that delivers a stabilized preset DC output via a probe
to an inspection electrode. As the electrode moves over
the coating surface, if it encounters a pin hole or
bare spot, a small current flows actuating a visible
(non-hazardous) spark at the point of contact and a
visible and audible alarm in the unit. |
 |
How is the applied voltage pre-set?
|
 |
The voltage required is pre-set manually on the unit
to a minimum level determined by the thickness of film
of the coating product and its generic type ie: PVC,
Teflon, FRP or polyethelene. Detailed instructions are
set out in the section of this handbook marked Recommended
Minimum Voltage. It is important to follow these instructions
in setting voltages as some surfaces have a much higher
dielectric strength than others - consequently offering
a high resistance to the conduction of electricity.
Applied test voltages should only be sufficient to detect
faults, otherwise overstressing of the dielectric strength
may occur with possible surface rupturing. |
 |
What are the minimum and maximum film thicknesses
that can be tested with Micromed Detectors?
|
 |
This depends on the type of coating applied. |
 |
Do damp coatings or moist and humid conditions affect
the operation of the Micromed Detector?
|
 |
As wet surfaces are generally conductive, this could
affect the unit operationally.
Atmospheric moisture is unlikely to do so.
|
 |
Is the high voltage application destructive to any
of the generally used sealing coatings?
|
 |
No! The applied voltage to the coating is non-destructive,
provided the voltage applied is within the parameters
set down in the Recommended Minimum Voltage section
of this handbook for the type and thickness of coating. |
 |
Can I get a shock while operating the unit?
|
 |
No! Under normal operating conditions the user should
not handling the ends of the ground wire and the electrode
wire at the same time. If this occurs then only a mild
tingle will be felt by the operator. |
If you have a question that is not covered in the FAQ
above email
us. |