MMPD 8K electrosurgical insulation defect detector

  • MMPD 8K
  • Optional Accessories
  • Specification
  • Support Documentation
  • Instruction Video
mm 5kv electrosurgical tester

McGan Technology’s MMPD-8K electrosurgical insulation defect detector is a battery operated unit that tests the integrity, such as pinholes, cracks or defects, of the insulation of electrosurgical instrument in order to prevent inadvertent tissue burns which may occur during electrosurgical instrument procedures. The integrity of the insulation of electrosurgical instruments can occur over time, through normal surgical use, through the accidental contact with a sharp instrument or the rubbing of the insulation to the point of abrading it. The insulation failure provides an alternate pathway for the current to leave the instrument's electrode usually beyond the sight of the surgeon which will result in the unintended tissue burn. Using an electrosurgical test equipment maintenance program can greatly reduce the number of these inadvertent burns during electrosurgical procedures.

Features:

  • Design incorporates a portable battery pack
  • Charges in 3 hours (overcharging protection)
  • Maintains applied test voltage with constant current source
  • Easy adjustable voltage control
  • Full test current at low voltages. Alarm sensitivity control.
  • Selection of test electrodes
  • Limited output current for operational safety
  • Regulated DC output, adjustable 0 to 8kV
  • Visual and audible alarms
  • LCD displays alarm and battery charge
  • Color coded cables and ports

Included accessories

Click on the thumbnails to see a larger photo or click on the manuals to download

Ground wire with alligator clip Ring and Brush Electrodes A/C Battery Charger
  • Ground Wire
  • Brush Electrode
 
  • A/C Battery Charger
 
HV Cable Carry Bag Quick Start Manual
  • HV Cable
 
  • Carry Bag
 
manual quick start
Brush Training CD
  • Brush
 
Training CD

Clink on any thumbnail below to see a larger photo

Wire Tester Unit

Tri-Hole Electrode

Saddle Block

  • Mcgan Wire Tester
 
  • Tri-Hole Electrode
 
  • Saddle
 
Compliance Specifications Typical Test Voltages
  • ANSI/AWWA C231-94
  • UL(power transformer)
  • CE marking in accordance with EEC directive 89/336
  • JIS G-3491, JIS G-3492
  • FDA registered (501K)
  • Weight: 4.85lbs (2.2kg)
  • Display: LCD
  • Voltage: 0V to 8kV adjustable
  • Resolution: 10V
  • Current output: <0.1mA at probe
  • Short circuit: test current 0.1mA max
  • Power supply: GellCell 1.5mA max
  • Dimensions: 10.25" x 6.3" x 2.75"
  • Alarm: audible and visual
  • Bipolar Micro: 0.4kV DC
  • Bipolar Standard: 2.5kV DC
  • Monopolar: 4.2kV DC

 

 


Take the first step to improving surgical safety and contact McGan Technology today.